Anterior segment optical coherence tomography angiography to evaluate the peripheral fitting of scleral contact lenses

Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/10045/78531
Información del item - Informació de l'item - Item information
Título: Anterior segment optical coherence tomography angiography to evaluate the peripheral fitting of scleral contact lenses
Autor/es: Gimenez-Sanchis, Inma | Palacios-Carmen, Beatriz | García-Garrigós, Angel | Cantó-Vañó, Javier | Pérez-Ortega, Antonio J. | Piñero, David P.
Grupo/s de investigación o GITE: Grupo de Óptica y Percepción Visual (GOPV)
Centro, Departamento o Servicio: Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía
Palabras clave: Scleral contact lens | Optical coherence tomography | OCT angiography | Apical clearance | Corneal limbus
Área/s de conocimiento: Óptica
Fecha de publicación: 4-sep-2018
Editor: Dove Press
Cita bibliográfica: Clinical Optometry. 2018, 10: 103-108. doi:10.2147/OPTO.S164454
Resumen: Purpose: The aim of this study was to show the potential applicability of optical coherence tomography angiography (OCTA) for the evaluation of the peripheral fitting of fully scleral contact lenses. Methods: A pilot study was proposed fitting three different scleral contact lenses (Irregular Corneal Design [ICD]) with different sagittal heights (4200, 4800, and 5600 mm) in a healthy volunteer of 27 years old. We evaluated by means of optical coherence tomography (OCT, DRI Triton) the apical clearance achieved with each of the three lenses fitted. The impact over scleral flow was assessed with the OCTA module of the same device. Results: The apical clearance was 310, 901, and 1680 µm with the scleral lenses of sagittal heights 4200, 4800, and 5600 µm, respectively. With OCTA, we evaluated the impact of the lens bearing on the conjunctival vascular flow, observing an area of vascular interruption of 0, 25, and 75% with the lenses of 4200, 4800, and 5600 µm of sagittal heights, respectively. The vascular interruption was induced in the perilimbar area, suggesting the need of readjusting the limbal clearance zone of the lens. Conclusion: Fully scleral contact lens fitting may be optimized with the use of OCTA, allowing the practitioner to perform the fitting with better control of the peripheral bearing of the lens on the conjunctival tissue, assessing the impact on vascular structures. This potential use of OCTA must be investigated further in future studies including large samples of eyes.
Patrocinador/es: The author DPP has been supported by the Ministry of Economy, Industry and Competitiveness of Spain within the program Ramón y Cajal, RYC-2016-20471.
URI: http://hdl.handle.net/10045/78531
ISSN: 1179-2752
DOI: 10.2147/OPTO.S164454
Idioma: eng
Tipo: info:eu-repo/semantics/article
Derechos: © 2018 Gimenez-Sanchis et al. This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
Revisión científica: si
Versión del editor: https://doi.org/10.2147/OPTO.S164454
Aparece en las colecciones:INV - GOPV - Artículos de Revistas

Archivos en este ítem:
Archivos en este ítem:
Archivo Descripción TamañoFormato 
Thumbnail2018_Gimenez-Sanchis_etal_ClinOptometry.pdf7,38 MBAdobe PDFAbrir Vista previa


Este ítem está licenciado bajo Licencia Creative Commons Creative Commons