Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability

Empreu sempre aquest identificador per citar o enllaçar aquest ítem http://hdl.handle.net/10045/61588
Información del item - Informació de l'item - Item information
Títol: Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
Autors: Piñero, David P. | López Navarro, Alberto | Cabezos, Inmaculada | Fez Saiz, Dolores de | Caballero, María T. | Camps, Vicente J.
Grups d'investigació o GITE: Grupo de Óptica y Percepción Visual (GOPV)
Centre, Departament o Servei: Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía
Paraules clau: Corneal | Topographic | Aberrometric | Multidiagnostic device | Healthy eyes
Àrees de coneixement: Óptica
Data de publicació: 2017
Editor: Hindawi Publishing Corporation
Citació bibliogràfica: Journal of Ophthalmology. Volume 2017 (2017), Article ID 2149145, 9 pages. doi:10.1155/2017/2149145
Resum: Purpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete anterior segment examination with the VX120 system (Visionix-Luneau Technologies, Chartres, France). Three consecutive measurements were obtained. The within-subject standard deviation (Sw), intrasubject precision (1.96xSw), and intraclass correlation coefficient (ICC) were calculated. Results. All Sw for corneal power measurements were below 0.26 D, with ICC above 0.982. The Sw for corneal astigmatism at different areas (3, 5, and 7 mm) was below 0.21 D, with ICC above 0.913. Concerning the axis of astigmatism, its Sw was below 11.27°, with ICC above 0.975. The Sw and ICC for corneal eccentricity were 0.067 and 0.957, respectively. The Sw and ICC for high-order aberration root mean square (RMS) were 0.048 µm and 0.901, respectively. For 3rd- and 4th-order aberrometric parameters, all Sw were below 0.037 µm and all ICC were higher than 0.84, except for quadrafoil RMS (ICC: 0.689). Conclusions. The multidiagnostic device evaluated is able to provide consistent measurements of corneal power, eccentricity, and third- and fourth-order aberrations in healthy eyes.
URI: http://hdl.handle.net/10045/61588
ISSN: 2090-004X (Print) | 2090-0058 (Online)
DOI: 10.1155/2017/2149145
Idioma: eng
Tipus: info:eu-repo/semantics/article
Drets: © 2017 David P. Piñero et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Revisió científica: si
Versió de l'editor: http://dx.doi.org/10.1155/2017/2149145
Apareix a la col·lecció: INV - GOPV - Artículos de Revistas

Arxius per aquest ítem:
Arxius per aquest ítem:
Arxiu Descripció Tamany Format  
Thumbnail2017_Pinero_etal_JOphtalm.pdf1,98 MBAdobe PDFObrir Vista prèvia


Aquest ítem està subjecte a una llicència de Creative Commons Llicència Creative Commons Creative Commons