Power line interference filtering on surface electromyography based on the stationary wavelet packet transform

Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/10045/39284
Información del item - Informació de l'item - Item information
Título: Power line interference filtering on surface electromyography based on the stationary wavelet packet transform
Autor/es: Galiana-Merino, Juan José | Ruiz-Fernandez, Daniel | Martínez Esplá, Juan José
Grupo/s de investigación o GITE: Sismología-Riesgo Sísmico y Procesado de la Señal en Fenómenos Naturales | Ingeniería Bioinspirada e Informática para la Salud
Centro, Departamento o Servicio: Universidad de Alicante. Departamento de Física, Ingeniería de Sistemas y Teoría de la Señal | Universidad de Alicante. Departamento de Tecnología Informática y Computación | Universidad de Alicante. Instituto Universitario de Física Aplicada a las Ciencias y las Tecnologías
Palabras clave: Electromyogram signal (EMG) | Biological signals filtering | Stationary wavelet packet transform (SWPT) | Power line interference
Área/s de conocimiento: Teoría de la Señal y Comunicaciones | Arquitectura y Tecnología de Computadores
Fecha de publicación: ago-2013
Editor: Elsevier
Cita bibliográfica: Computer Methods and Programs in Biomedicine. 2013, 111(2): 338-346. doi:10.1016/j.cmpb.2013.04.022
Resumen: Power line interference is one of the main problems in surface electromyogram signals (EMG) analysis. In this work, a new method based on the stationary wavelet packet transform is proposed to estimate and remove this kind of noise from EMG data records. The performance has been quantitatively evaluated with synthetic noisy signals, obtaining good results independently from the signal to noise ratio (SNR). For the analyzed cases, the obtained results show that the correlation coefficient is around 0.99, the energy respecting to the pure EMG signal is 98–104%, the SNR is between 16.64 and 20.40 dB and the mean absolute error (MAE) is in the range of −69.02 and −65.31 dB. It has been also applied on 18 real EMG signals, evaluating the percentage of energy respecting to the noisy signals. The proposed method adjusts the reduction level to the amplitude of each harmonic present in the analyzed noisy signals (synthetic and real), reducing the harmonics with no alteration of the desired signal.
URI: http://hdl.handle.net/10045/39284
ISSN: 0169-2607 (Print) | 1872-7565 (Online)
DOI: 10.1016/j.cmpb.2013.04.022
Idioma: eng
Tipo: info:eu-repo/semantics/article
Revisión científica: si
Versión del editor: http://dx.doi.org/10.1016/j.cmpb.2013.04.022
Aparece en las colecciones:INV - IBIS - Artículos de Revistas
INV - GIRS - Artículos de Revistas

Archivos en este ítem:
Archivos en este ítem:
Archivo Descripción TamañoFormato 
Thumbnail2013_Galiana_etal_CMPB_final.pdfVersión final (acceso restringido)1,24 MBAdobe PDFAbrir    Solicitar una copia


Todos los documentos en RUA están protegidos por derechos de autor. Algunos derechos reservados.