New theoretical matrix formula for intraocular lens calculation using the optimal bending factor
Empreu sempre aquest identificador per citar o enllaçar aquest ítem
http://hdl.handle.net/10045/9270
Registre complet
Camp Dublin Core | Valor | Idioma |
---|---|---|
dc.contributor | Holografía y Procesado Óptico | en |
dc.contributor.author | Fimia Gil, Antonio | - |
dc.contributor.author | Alió y Sanz, Jorge L. | - |
dc.contributor.author | Pascual, Inmaculada | - |
dc.contributor.author | Beléndez, Augusto | - |
dc.contributor.other | Universidad de Alicante. Departamento de Óptica, Farmacología y Anatomía | en |
dc.contributor.other | Universidad de Alicante. Departamento de Física, Ingeniería de Sistemas y Teoría de la Señal | en |
dc.date.accessioned | 2009-01-29T11:52:47Z | - |
dc.date.available | 2009-01-29T11:52:47Z | - |
dc.date.created | 1992 | - |
dc.date.issued | 1993-03 | - |
dc.identifier.citation | FIMIA GIL, Antonio, et al. "New theoretical matrix formula for intraocular lens calculation using the optimal bending factor". Journal of Cataract & Refractive Surgery. Vol. 19 (March 1993). ISSN 0886-3350, pp. 293-297 | en |
dc.identifier.issn | 0886-3350 (Print) | - |
dc.identifier.issn | 1873-4502 (Online) | - |
dc.identifier.uri | http://hdl.handle.net/10045/9270 | - |
dc.description.abstract | By applying Gaussian optics we propose a new matrix formula model for intraocular lens (IOL) calculation that yields not only IOL optical power for a particular eye but also an evaluation of the bending factor of different IOL optical designs. This option improves IOL selection and results in the best possible optical image, a function dependent on the IOL's bending factor. This formula is a new approach to IOL calculation and may be useful in designing IOLs for abnormally hyperopic or myopic eyes. Optimal IOL power and design should both be considered in IOL calculation, which may improve the optical and clinical results of IOL implantation in the aphakic patient. The radius of the cornea is assumed to be spherical, and aspherical curves of the cornea may influence the optimal "bending" or shape factor of a lens. | en |
dc.language | eng | en |
dc.publisher | American Society of Cataract and Refractive Surgery | en |
dc.publisher | European Society of Cataract and Refractive Surgeons | en |
dc.publisher | Elsevier | en |
dc.subject | Aphakia | en |
dc.subject | Bending factor | en |
dc.subject | Intraocular lens | en |
dc.subject | Optical aberration | en |
dc.subject | Optical power | en |
dc.subject | Matrix calculation | en |
dc.subject | Retinal image | en |
dc.subject.other | Óptica | en |
dc.title | New theoretical matrix formula for intraocular lens calculation using the optimal bending factor | en |
dc.type | info:eu-repo/semantics/article | en |
dc.peerreviewed | si | en |
dc.identifier.doi | 10.1016/S0886-3350(13)80959-5 | - |
dc.relation.publisherversion | http://dx.doi.org/10.1016/S0886-3350(13)80959-5 | - |
dc.rights.accessRights | info:eu-repo/semantics/restrictedAccess | - |
Apareix a la col·lecció: | INV - GHPO - Artículos de Revistas |
Arxius per aquest ítem:
Arxiu | Descripció | Tamany | Format | |
---|---|---|---|---|
JCRS_v19_n2_p293_1993.pdf | Versión final (acceso restringido) | 913,66 kB | Adobe PDF | Obrir Sol·licitar una còpia |
JCRS_v19_n2_p293_1993pre.pdf | Versión revisada (acceso libre) | 1,05 MB | Adobe PDF | Obrir Vista prèvia |
Tots els documents dipositats a RUA estan protegits per drets d'autors. Alguns drets reservats.